  1. Make sure the O-rings on the upper and lower tubes are in good shape. To check that, remove the O-rings by removing the cap nut on the bottom and top of the reactor. Carefully insert the glass tube and gently remove the O-ring by moving the glass tube sideways by 5-10° and then pulling the O-ring out. Check for tears and dirt on the O-ring and replace the O-ring if necessary. If the O-ring still looks intact but very dry, re-grease the O-ring with a very small amount of silicone grease from the EA-kit.

  2. Reassemble the system and test again according to section 1 of the ‘Diagnosis’run another diagnosis.

  3. If the problem is till there, the leak will likely be found at the joint of the reactor housing with the 1/16” capillary. In this case, apply some epoxy glue around the joint of the reactor and the capillary.

  4. Test again according to section 1 of the ‘Diagnosis’Run another diagnosis.

  5. If your the leak is still therepersists, contact Ionplus support.Support

Fixing a leak through valve:

...