Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

\uD83E\uDD14 Problem

Pressure increases at underpressure (decreases at overpressure) over time

This indicates a leak

\uD83C\uDF31 Solution

Diagnosis - Determine the type of leak (atmosphere or valve)

Note

Make sure no samples are in the AGE system and insert a glass tube on each reactor of the AGE system.

...

  1. Make sure the O-rings on the upper and lower tubes are in good shape. To check that, remove the or O-rings by removing the cap nut on the bottom and top of the reactor. Carefully insert the glass tube and gently remove the O-ring by moving the glass tube sideways by 5-10° and then pulling the O-ring out. Check for tears and dirt on the O-ring and replace the O-ring if necessary. If the O-ring still looks intact but very dry, regrease re-grease the O-ring with a very small amount of silicone grease from the EA-kit.

  2. Reassemble the system and test again according to section 1 of the ‘Diagnosis’run another diagnosis.

  3. If the problem is till there, the leak will likely be found at the joint of the reactor housing with the 1/16” capillary. In this case, apply some epoxy glue around the joint of the reactor and the capillary.

  4. Test again according to section 1 of the ‘Diagnosis’Run another diagnosis.

  5. If your the leak is still therepersists, contact Ionplus support.Support

Fixing a leak through valve:

...

  1. If the leak is in the valve, we recommend you get in touch with Ionplus support and likely the valve will need to be replaced.

...

  1. contact Ionplus Support to request a replacement valve.

...