Skip to end of metadata
Go to start of metadata

You are viewing an old version of this page. View the current version.

Compare with Current View Page History

« Previous Version 3 Next »

First of all, remember to perform the following good practices to prevent clogging of the needle at all times:

  • Replace the septa of the «Flushing» position after each batch of samples.

  • Soak the needle in some milli-Q water after each use in order to prevent potential damage at the needle due to long-term contact with the acid.

  • Clean potential acid drops on every septa after acid addition.

🤔 Problem

The needle shows no or smaller Helium flow, indicating that the needle may be clogged.

🌱 Solution

  • Check visually under a microscope for potential waste blocking the needle tip- and sideholes and carefully remove it.

  • If no waste is visible, try any of the following actions:

    • soak the needle in some milli-Q water and flush the needle with Helium to dry it. If the blocking persists, soaking in ethanol instead can be helpful, though extra thorough drying should be then performed

    • try to carefully clean the two holes in the double-walled headspace needle with a thin wire

    • flush the needle under higher pressure (eg. 3 mbar) in order to remove potential remnants of septa or other waste inside the needle

    • switch the capillaries, connecting the sample-out capillary to the sample-in at the back of the CHS2 to flush out particles from the side hole, applying either normal or higher (eg. 3 mbar) pressure

    • bend the tip of the needle over the capillary (it should be done at Ionplus)

 

If none of the aforementioned actions help, swapping-out the needle might be probably the most reasonable solution!

  • No labels