No registered 14C counts
Issue Description
There are no registered 14C counts.
Problem Solving
Start a single run on a known standard (OxII).
Check that the 12C currents are still present (20 - 30 μA 12C+ on a solid sample).
Look up tuning settings from a previous time when the system was working:
Compare with present settings to see if a setting is significantly different than usual.
Apply the old settings.
If the 14C counts are still missing, check the detector settings:
Check that the bias voltage is set.
Check that the detector pressure is stable.
Adjust the discriminator setting if necessary.
Make sure the slits are not blocking the beam.